About ISO Standards 10360
|
|
 |
MPEE
Volumetric Length measuring error MPEE
A set of 5 gages has to be measured 3 times with one probing at each
end, while the block is being aligned in 7 different directions in space.
All measuring results must fulfil the Specification "E".
|
|
 |
MPEP
Volumetric probing Error MPEP
A Precision sphere has to be measured with 25 equally distributed
probings. P is the range of all radii (sphere form). (Rmax
- Rmin = Sphere form).
|
|
 |
MPETHP
Volumetric Scanning probing Error MPETHP
A precision Sphere is to be scanned with 4 defined lines. THP is the
range of all radii (sphere form-scanning) (Rmax -
Rmin = Sphere form)
|
|
 |
MF, MS, ML
Fixed multiple-stylus probing system error MF, MS, ML
A precision sphere is to be measured with a 5 probe-star (fixed probe
head) or with 1 probe in 5 different orientations (revolving probe head),
with 5x25 probing. Form, size and location error over all 125 probings. |
|